Product Parameters
XY-560B1 Anesthesia machine(The standard)
Generic evaporator, either Halothane, Enflurane, Isoflurane or Sevoflurane configuration. Double-tubes O2flowmeter, standard loopsystem, ABS engineered plastic frame.
Part of Anesthesia
Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
Scope of application adult, child
Gas source O 2 : 0.28~0.6MPa
Flow meter O 2 : 0.05~1.0L/min 1.1~10L/min
Flow rate of rapid oxygen supply 25~75L/min
Low oxygen pressure alarm There will be sound alarm when the oxygen
pressure < 0.2MPa
Vaporizer it has the function of automatic compensation
based on pressure, temperature, and flow rate.
The regulation range of evaporator
concentration is 0~5 vol%.
Among Halothane, Enflurane, Isoflurane and
Sevoflurane, one can be chosen for
application as required by the customer.
Respiratory circuit working mode: all-close, semi-close, semi-open
APL:≤12.5 kPa
Respiratory bellows bellows for adults, bellows for children
tidal volume range:0~1500 ml
Part of Ventilator
Display mode High-definition 5.7" LCD screen display
Ventilation mode IPPV, SIPPV, SIMV, MANUAL
Ventilation parameters Tidal volume 50~1500ml
Rate 2~99 bpm
SIMV 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure -10~20cmH 2 O
PEEP 2~20cmH 2 O
Pressure limitation 10~60cmH 2 O
SIGH 1.5 times the inspiratory time/100bpm
Parameters for ventilation monitoring tidal volume, minute ventilation volume,
IPPV rate, SIMV rate, total respiratory rate,
I/E, peak pressure of airway, pressure - time
waveform, flow - time waveform, PEEP,
inspiratory trigger pressure
Security Alarm System
Airway pressure alarm upper limit setting range 5~60 cmH 2 O
low limit setting range 0~50 cmH 2 O
Minute ventilation volume alarm upper limit setting range 1~20 L/min
low limit setting range 0~19 L/min
Tidal volume upper limit setting range 100~1500ml
low limit setting range 0~1400ml
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 25 cmH 2 O
(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)
Power AC 220V 50Hz (UPS, Storage battery)
(Mechanical Arms) (End-tidal C O 2 monitoring)
(Specific interface of waste anesthesia gas)
*Items within brackets are optional.